VITALITY DS (model T125) and VITALITY 2 (models T165, 175) ICD (implantable cardiac defibrillator)

FDA device recall Z-0892-06 · posted 2006-05-23 · Terminated

Recall details

Recalling firm
Guidant Corporation
Reason for recall
Devices manufactured with batteries built on April 26 or 27, 2005, by a single operator, may have battery welds that may have been performed at settings outside of typical manufacturing specifications at the supplier.
Action taken
Guidant issued a Physicain letter, dated 10/25/05 to implanting/following physicians. The letter states the issues and reommends patient follow-up care every three months, as stated in published labeling. An Advisory Update, dated 12/20/05, was sent to physicians that describes the issue, states that here have been zero reports of battery weld failures on implanted devices and the results of laboratory testing (accelerated testing) in which they tried to induce the failure. The results of the testing induced 2 battery weld post failures. A projected rate of occurence was also given. Patient management recommendations have not changed.
Root cause
Other
Status
Terminated
Product code
LWP
Quantity affected
40
Distribution
Nationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland,
Date initiated
2005-09-30
Date posted
2006-05-23
Date terminated
2007-01-16
Location
Saint Paul, MN