VITALITY AVT, model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Guidant Corporation
FDA device recall Z-1291-06 · posted 2006-07-27 · Terminated
Recall details
- Recalling firm
- Boston Scientific CRM Corp
- Reason for recall
- Guidant has identified a pattern of premature battery depletion in devices manufactured with a single lot of capacitors from a single supplier. If function of this low voltage capacitor is compromised, the battery may deplete prematurely.
- Action taken
- An Urgent Medical Device Safety Information & Corrective Action letter, dated May 12, 2006, was sent to physicians. The letter gave a description of the issue, clinical implications, rate of occurrence, devices affected and recommendation. Physicians are asked to follow-up with affected patients as soon as possible because ERI to EOL time may be shortened. Physicians are asked to contact Guidant for instructions on performing a baseline 'save to disk' and provide this information to Guidant. Guidant will be able to analyze the device memory data and estimate remaining longevity and provide individualized follow-up and replacement guidelines. Subsequent follow-up visits should continue to monitor for evidence of premature battery depletion.
- Root cause
- Other
- Status
- Terminated
- Product code
- LWP
- Quantity affected
- 66
- Distribution
- Nationwide including Puerto Rico. Worldwide, including Argentina, Belgium, Canada, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, New Zealand, Portugal, Spain, Syrian Arab Republic, Un
- Date initiated
- 2006-05-15
- Date posted
- 2006-07-27
- Date terminated
- 2008-03-02
- Location
- Saint Paul, MN
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.