Vital Images ViTALConnect , Version 4.0 and 4.1, PET/CT Visualization option. Vital Images, Inc., 5850 Opus Parkway, Sui

FDA device recall Z-0520-2008 · posted 2008-01-31 · Terminated

Recall details

Recalling firm
Vital Images, Inc.
Reason for recall
Standard Uptake Values (SUV) display/update problems; in the way Standard Uptake Values (SUV) are displayed in 3D/MPR views. The problem arises when a user draws a selected region of interest in order to calculate the SUV statistics of the region. If the user then changes the slice shown by scrolling through additional slices, the original SUV measurements will remain on the screen and do not up
Action taken
Customers were notified via telephone (11/7 -11/8/2007) and issued an Urgent Medical Device recall notification letter on 11/8/07. The letter described the problem and product involved. The letter also informed customers they they will provide a free software patch within the next 6-8 weeks.
Root cause
Process control
Status
Terminated
Product code
LLZ
Quantity affected
18
Distribution
Worldwide-USA including states of SC, MO, MI, WI, NY, FL, CO, CA, PA, and WA, and countries of England and The Netherlands
Date initiated
2007-11-08
Date posted
2008-01-31
Date terminated
2011-12-17
Location
Plymouth, MN