VistaSolution, physiological signal transmitter. A software graphical user interface for use by healthcare professionals
FDA device recall Z-2537-2020 · posted 2020-07-10 · Terminated
Recall details
- Recalling firm
- Vitalconnect Inc.
- Reason for recall
- A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification. The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.
- Action taken
- On May 21, 2020, the firm sent a "Urgent: Medical Device Recall" notification to all affected customers via E-mail and mail. The letter stated, that the third party Sp02 continuous device notifications may not indicate an alert as intended. Therefore, the health care provider may not be notified to check the patient. Potential Hazards: A patient may have hypoxia and the designated health care provider may not be notified. The firm will be turning off the ability to set custom notifications containing continuous SP02 until the issue is resolved. The firm tells customers to monitor the patients for SPO2, as if the patients are not connected to the recalled device. The customers are informed to acknowledge the receipt of this notification by E-mail or via mail to the firm, email to: cmerrell@vitalconnect.com and if any questions, then contact: Tony Wilder at phone: (408) 963-4600 or support@vitalconnect.com
- Root cause
- Software Design Change
- Status
- Terminated
- Product code
- DRG
- Quantity affected
- 7
- Distribution
- U.S. Nationwide distribution including in the states of NJ, NY, VA, OUS: None
- Date initiated
- 2020-05-21
- Date posted
- 2020-07-10
- Date terminated
- 2021-05-13
- Location
- San Jose, CA