Vista Dental Products CanalClean Endodontic Irrigation Kit, (1) 12mL - Chlor-XTRA, 6mL - SmearOFF + 92) + (2) 30ga Irrig
FDA device recall Z-0011-2020 · posted 2019-10-01 · Terminated
Recall details
- Recalling firm
- Inter-Med Llc
- Reason for recall
- The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
- Action taken
- The firm initiated the recall by letter on 06/27/2019. The letter directed the consignee to quarantine the product and contact the returns department for return instructions.
- Root cause
- Process control
- Status
- Terminated
- Product code
- KJJ
- Quantity affected
- 377 units
- Distribution
- US, South Africa, Switzerland, Poland, Italy, Singapore, Australia
- Date initiated
- 2019-06-27
- Date posted
- 2019-10-01
- Date terminated
- 2021-08-10
- Location
- Racine, WI