ViewPoint; system, imaging processing, radiological
FDA device recall Z-3145-2024 · posted 2024-09-13 · Open, Classified
Recall details
- Recalling firm
- GE Healthcare GmbH
- Reason for recall
- The National Institute of Standards and Technology (NIST) has identified in the National Vulnerability Database (NVD) that NexGen Healthcare Mirth Connect, prior to version 4.4.1, has a potential vulnerability allowing the deserialization of untrusted data (CVE-2023-43208). Mirth products may have been installed and configured with your ViewPoint / ViewPoint 6 system by GE HealthCare. This vulnerability could allow a malicious actor to access the system and potentially manipulate patient data.
- Action taken
- GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice on 08/08/2024 via US Mail. The notice explained the issue, potential risk, and requested the following: You can continue to use your device by following the instructions below: 1.Ensure ViewPoint / ViewPoint 6 is operating in a secure hospital Network such that: a. only authorized personnel have access, b. protections against denial of service and wiretapping/network sniffing are installed by the network operators, and c. access from the Internet is limited where possible 2.Ensure ViewPoint / ViewPoint 6 is operating in a secure Windows environment including confirmation that all relevant software is up to date and operating systems updates are installed. Please ensure all potential users in your facility are made aware of this safety notification.
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- LLZ
- Quantity affected
- 3 units
- Distribution
- US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana,
- Date initiated
- 2024-08-08
- Date posted
- 2024-09-13
- Location
- Munich, Germany
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| GE VIEWPOINT (K050943) | General Electric Co. | 2005-05-23 |