Victory 9 with Power Seat (SC609PS) Motorized wheelchair with elevating seat.
FDA device recall Z-1941-2012 · posted 2012-07-03 · Terminated
Recall details
- Recalling firm
- Pride Mobility Products Corp
- Reason for recall
- The main harness connector was manufactured without the required back shield sealing protection. As a result, there is the potential for water to seep through the nonconforming harness.
- Action taken
- Three (3) pieces of information was distributed in the notification. They are: (1) Letter of Correction to the Provider (2) Letter of Correction for the Consumer (provider to issue to consumer) (3) Product Insert for Victory 9 Controller Interface Harness Replacement. The letter of Correction for Victory 9 with Power Sear (SC609PS), dated November 2010, identified the affected product and the reason for the correction. Customers were to record the serial number(s) and contact their provider's office to make an appointment to have the harness replaced.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- INI
- Quantity affected
- 186
- Distribution
- Worldwide Distribution
- Date initiated
- 2010-11-26
- Date posted
- 2012-07-03
- Date terminated
- 2013-03-22
- Location
- Exeter, PA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| VICTORY THREE WHEEL SCOOTER, MODEL: SC610 (K072165) | Pride Mobility Products Corp. | 2007-08-17 |