Viasys Healthcare NicoletOne ICU Monitor, Modular Neurodiagnostic System (containing Axiomtek Panel PC computer system),

FDA device recall Z-0126-2008 · posted 2008-01-03 · Terminated

Recall details

Recalling firm
Nicolet Biomedical Div of Viasys Healthcare
Reason for recall
Display lock-up: VIASYS Healthcare NicoletOne ICU Monitor freezes up during operation when using the digital video option.
Action taken
Certified direct mail delivery of URGENT MEDICAL DEVICE FIELD CORRECTION (RECALL) were made to consignees on 9/11/07 by Nicolet Biomedical Div of Viasys Healthcare. The letter describes the product and problem. VIASYS requests that customers cease use of the video option of the device until the monitor can be replaced by VIASYS authorized personnel. Letter also requests consignees to fax or mail back the response form.
Root cause
Component design/selection
Status
Terminated
Product code
GWQ
Quantity affected
33
Distribution
Worldwide Dsitribution - USA, Germany, Netherlands, Kuwait, Japan, Ireland, and China.
Date initiated
2007-09-11
Date posted
2008-01-03
Date terminated
2011-12-10
Location
Madison, WI