Atlas FDA

Version 12 software program written for the Bio-Tek ELx800 automated microtiter plate reader with use of PRO-Trac Tacrol

FDA device recall Z-0702-05 · posted 2005-04-13 · Terminated

Recall details

Recalling firm
Diasorin Inc.
Reason for recall
The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert.
Action taken
Customers were sent notification letters and effectiveness check form. recommended users manually veify results using product insert specifications.
Root cause
Other
Status
Terminated
Product code
MLM
Distribution
CA, Washington DC, FL, KY & Canada
Date initiated
2004-04-12
Date posted
2005-04-13
Date terminated
2006-07-22
Location
Stillwater, MN

Monitor recalls programmatically