VersaCross Access Solution (VXSK), Product Description and ID: VXSK (180J RF WIRE 63CM/45DEG/D0) VXSK0101 VXSK (180J RF

FDA device recall Z-2103-2025 · posted 2025-07-10 · Open, Classified

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Action taken
On June 10, 2025, the firm began notifying consignees of the field action via Urgent Medical Device Removal letters. Customers are instructed to not use affected product. They should remove affected units from their inventory and segregate the units in a secure place until they can be returned to Boston Scientific.
Root cause
Packaging change control
Status
Open, Classified
Product code
DXF
Quantity affected
14,724
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.
Date initiated
2025-06-10
Date posted
2025-07-10
Location
Saint Paul, MN