Veritas Advanced Infusion Packs, REF: VRT-AI

FDA device recall Z-1005-2023 · posted 2023-01-20 · Open, Classified

Recall details

Recalling firm
Johnson & Johnson Surgical Vision, Inc.
Reason for recall
Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
Action taken
On 12/07/22, recall notices were sent to customers who were asked to do the following: 1) Discontinue using affected product. 2) Return product to the recalling firm. 3) Complete and return the Customer Reply form via email to RegCompliOne@its.jnj.com 4) Distributors, distribute the recall notice to your customers. Customers with questions or concerns can call the firm at 1-877-266-4543 Option 2.
Root cause
Process control
Status
Open, Classified
Product code
HQC
Quantity affected
3133
Distribution
US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom
Date initiated
2022-12-07
Date posted
2023-01-20
Location
Irvine, CA