Atlas FDA

Ventri Discovery NM 530c, Ventri with Premium Table Discovery NM 530c, models H3000ZW & H3000SA. The system is used to p

FDA device recall Z-1168-2013 · posted 2013-04-24 · Terminated

Recall details

Recalling firm
GE Healthcare, LLC
Reason for recall
GE Healthcare has recently become aware of an incident which caused a patient injury during an unload following an exam on a Ventri scanner. During a patient unload following a scan, the patients fingers may be pinched between the moving cradle edge and the stationary base of the table while the table is moving outward automatically.
Action taken
Consignees were sent on 1/9/2013 a GE Healthcare "Urgent Medical Device Correction" letter dated January 9, 2013. The letter was addressed to Managers of Nuclear Medicine, Hospital Administrators / Risk Managers and Managers of Radiology / Head of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information.
Root cause
Device Design
Status
Terminated
Product code
KPS
Quantity affected
556
Distribution
Worldwide Distribution - US, including the states of AL, AZ, AR, CA, CO, CT, DE, DC, FL, IL, IN, IA, KY, LA, ME, MD, MA, MI, ,N,MS, MO, NE, NV, NH, NJ, NM, NY,NC, OH, OK,OR, PA, RI, SC, TN, TX, UT, VA
Date initiated
2013-01-09
Date posted
2013-04-24
Date terminated
2013-08-06
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
VENTRI 1.1 (K080124)Ge Medical Systems F.I. Haifa2008-01-31
VENTRI (K051855)Ge Medical Systems F.I. Haifa2005-08-02