Vascular Solutions, Pronto V4, Extraction Catheter, Sterilized using ethylene oxide, Rx Only. The following sizes/model

FDA device recall Z-2946-2011 · posted 2011-08-04 · Terminated

Recall details

Recalling firm
Vascular Solutions, Inc.
Reason for recall
Vascular Solutions , Inc recently discovered that there was a manufacturing defect with Pronto V4 extraction catheters (Models 4005, 4006, 4007, 4008). Use of this defective product could lead to a tip separation. This may cause injury and/or procedural delay, resulting in a deteriorating hemodynamic condition of the patient which may require medical intervention.
Action taken
Vascular Solutions, Inc. sent an "URGENT: MEDICAL DEVICE FIELD ACTION" letter dated July 8, 2011 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The letter instructs customers to identify the affected product and remove from current inventory. Included with the letter was a Field Action Customer Inventory Form for customers to complete and return. VSI will contact customers for replacement product upon receipt of the form and returned product. Contact your local representative for questions concerning this notice.
Root cause
Process control
Status
Terminated
Product code
DXE
Quantity affected
1086
Distribution
Nationwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DC, FL, ID, IL, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK ,OR, PA, SC, SD, TN, TX, UT,
Date initiated
2011-07-11
Date posted
2011-08-04
Date terminated
2012-04-03
Location
Maple Grove, MN