Atlas FDA

VariSource CT/MR Ring & Tandem Applicator Set, catalog # AL13017000, containing 32mm titanium ring applicators, each ide

FDA device recall Z-2081-2008 · posted 2008-09-16 · Terminated

Recall details

Recalling firm
Varian Medical Systems, Inc.
Reason for recall
Medical device may exhibit inaccurate positioning and lead to unintended dose delivery during brachytherapy treatment.
Action taken
The firm notified medical device users by letter on 03/12/08. The notification provided updated instructions for use for the CT/MR Ring and Tandem Combination Applicator Set. Users were advised to characterize all ring applicators to develop an offset correction for each dwell position in order to construct a plan to deliver the dwell positions accurately and reliably. The characterization can be done by creating a set of autoradiographs with intended dwell positions and define treatment plan shifts in order to deliver dwell positions in intended locations. Users were advised that a subsequent set of verification autoradiographs should be made using the derived plan shifts to verify the corrected source positioning. The notification also advised that an individual ring's behavior can change over its duration of use and that periodic commissioning is dictated to ensure reproducibility and compensate for changes in the device over time. Users were instructed to inform all employees and extend the notice to any consignees. Users were requested complete and return the enclosed response form. If you have questions about this recall, send an email to brachyhelp@varian,com.
Root cause
Device Design
Status
Terminated
Product code
JAQ
Quantity affected
120 units
Distribution
The medical devices were distributed to radiation oncology facilities nationwide.
Date initiated
2008-03-12
Date posted
2008-09-16
Date terminated
2012-11-28
Location
Charlottesville, VA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CT HDR TANDEM/RING APPLICATOR WITH RECTAL RETRACTOR (K030110)Mick Radio-Nuclear Instruments, Inc.2003-04-03