Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient p

FDA device recall Z-0526-2015 · posted 2014-12-16 · Terminated

Recall details

Recalling firm
Varian Medical Systems, Inc.
Reason for recall
Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.
Action taken
Varian will work with affected customers to establish schedules that minimize impact to their clinic's treatment schedule and send Field Service Engineers with the required hardware to each affected site to make the modification.
Root cause
Component change control
Status
Terminated
Product code
IYE
Quantity affected
32 units
Distribution
Worldwide Distribution.
Date initiated
2014-04-30
Date posted
2014-12-16
Date terminated
2015-03-16
Location
Palo Alto, CA