Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Prod

FDA device recall Z-1325-2013 · posted 2013-05-15 · Terminated

Recall details

Recalling firm
Varian Medical Systems, Inc. Oncology Systems
Reason for recall
Under certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may result in no more than about 3.5% dose deviation from expected at any point in the beam, and may result in minor injury to the patient or slightly decreased local disease control.
Action taken
Varian sent a Customer Technical Bulletin to all affected customers. The bulletin identified the affected product, problem and actions to be taken. For questions, issues, or concerns contact Varian at 1-888-827-4265 or go to http://my.varian.com
Root cause
Process design
Status
Terminated
Product code
IYE
Quantity affected
3943 units
Distribution
Worldwide distribution - USA (nationwide) including Puerto Rico, and countries of: Algeria, Angola, Australia, Austria, Bahrain, Bangladesh, Belarus, Belguim, Brazil, Bosnia and Hezregovina, Bulgaria,
Date initiated
2011-05-31
Date posted
2013-05-15
Date terminated
2013-05-15
Location
Palo Alto, CA