Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algori
FDA device recall Z-0867-2013 · posted 2013-02-20 · Terminated
Recall details
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems
- Reason for recall
- The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.
- Action taken
- Varian Medical Systems, Inc. Oncology Systems sent an Urgent Medical Device Correction Letter dated January 30, 2013 to all affected customers. The Letter identified the affected product, problem and actions to be taken. Varian Medical Systems is developing a technical correction for this issue, Its customers will be contacted by a Varian Customer Service representatives when this correction is available to schedule its installation. The letter provides customers with help desk contact information.
- Root cause
- Software design
- Status
- Terminated
- Product code
- MUJ
- Quantity affected
- 1168 units
- Distribution
- Worldwide Distribution - USA Nationwide
- Date initiated
- 2013-01-28
- Date posted
- 2013-02-20
- Date terminated
- 2014-04-04
- Location
- Palo Alto, CA