Atlas FDA

Varian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algori

FDA device recall Z-0867-2013 · posted 2013-02-20 · Terminated

Recall details

Recalling firm
Varian Medical Systems, Inc. Oncology Systems
Reason for recall
The calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.
Action taken
Varian Medical Systems, Inc. Oncology Systems sent an Urgent Medical Device Correction Letter dated January 30, 2013 to all affected customers. The Letter identified the affected product, problem and actions to be taken. Varian Medical Systems is developing a technical correction for this issue, Its customers will be contacted by a Varian Customer Service representatives when this correction is available to schedule its installation. The letter provides customers with help desk contact information.
Root cause
Software design
Status
Terminated
Product code
MUJ
Quantity affected
1168 units
Distribution
Worldwide Distribution - USA Nationwide
Date initiated
2013-01-28
Date posted
2013-02-20
Date terminated
2014-04-04
Location
Palo Alto, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ECLIPSE TREATMENT PLANNING SYSTEM (K102011)Varian Medical Systems, Inc.2010-09-03