VARI-STIM III Nerve Locator/Stimulator, Sterile, Medtronic Xomed, Inc. The Vari-Stim III nerve locator/stimulator is a s

FDA device recall Z-0947-2009 · posted 2009-03-03 · Terminated

Recall details

Recalling firm
Medtronic Xomed, Inc.
Reason for recall
Device may fail to operate or operate intermittently during use and such operation could result in a lack of nerve stimulation when nerve tissue is actually contacted by the stimulator.
Action taken
All consignees were contacted by telephone (initiated on November 10, 2008). After telephone contact, each consignee was sent a follow-up fax/letter, "Urgent Product Recall Notification". The letter described the issue and actions for customers to take in response to the recall. Customers were instructed to check their inventories and return affected product to Medtronic Xomed, Inc. Customers were also instructed to fax their completed "Customer Contact and Product Return Checklist" to Medtronic Xomed, Inc. at 1-904-296-2386. Direct questions to Medtronic ENT Customer Service at 1-800-874-5797 (select Option 1 for Customer Service) and ask for the Vari-Stim Field Action Contact Person.
Root cause
Employee error
Status
Terminated
Product code
ETN
Quantity affected
22 boxes (220 units)
Distribution
Worldwide Distribution --- including USA and country of Canada. (22 boxes were distributed within U.S. Three boxes were distributed to Medtronic Canada but returned)
Date initiated
2008-11-21
Date posted
2009-03-03
Date terminated
2009-10-01
Location
Jacksonville, FL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
STIMULUS/DISSECTION INSTRUMENTS, BALL-TIP PROBES (K031003)Medtronic Xomed, Inc.2003-04-11