Vapofils with Stainless Steel Block Ends, Item Number V0507F (Isoflurane). The vapofil is a keyed anesthetic transfer tu

FDA device recall Z-1189-06 · posted 2006-07-04 · Terminated

Recall details

Recalling firm
Southmedic Inc
Reason for recall
Southmedic has initiated this recall because their Japanese distributor received field reports of leaking vapofils. Leakage may cause non-toxic spills in the hospital environment, distraction of practitioners during conduct of anesthesia and surgery, and may damage certain plastic equipment. Leakage may also expose patients to low concentrations of volatile anesthetics.
Action taken
Recall letters were sent to both US consignees via US Postal Service delivery with signature receipt required on April 21, 2006. Each letter included a Recall and Effectiveness Check Form for each consignee to fill out and fax to Southmedic. U.S. and international consignees were instructed to examine inventory immediately to identify if they have any product on hand, discontinue the distribution of these products and return all affected products to Southmedic. Consignees were also instructed to immediately notify their customers if they have further distributed these products.
Root cause
Other
Status
Terminated
Product code
CAD
Quantity affected
69 in US (211 Worldwide)
Distribution
US in FL & NJ Worldwide in Australia, Belgium, Ecuador, Germany, Japan, Netherlands, Sri Lanka, Taiwan, & United Arab Emerates
Date initiated
2006-04-21
Date posted
2006-07-04
Date terminated
2012-03-06
Location
Barrie Canada, Ontario