VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

FDA device recall Z-0327-2026 · posted 2025-10-24 · Open, Classified

Recall details

Recalling firm
Canon Medical System, USA, INC.
Reason for recall
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Action taken
Cannon Medical began notifying consignees via email on about 10/16/2025. Consignees were instructed that a representative of Canon Medical Systems USA will contact them to schedule a time to replace the relevant parts. If the message "Gradient power supply error is displayed before the replacement is performed, immediately stop using the system and contact your service representative. Consignees were requested to complete and return the provided response form
Root cause
Device Design
Status
Open, Classified
Product code
LNH
Quantity affected
2 units
Distribution
US distribution to Ohio and Nevada
Date initiated
2025-10-16
Date posted
2025-10-24
Location
Tustin, CA