Atlas FDA

Valeant Pharmaceuticals North America, Small Particle Size Aerosol Generator, SPAG-2 6000 series. Model No. SPAG-2. A ne

FDA device recall Z-0674-2016 · posted 2016-01-21 · Terminated

Recall details

Recalling firm
Valeant Pharmacueticals International
Reason for recall
The Serial I.D. Label and some information in the Operation Manual for the SPAG 2 6000 series is incorrect or out of date, including references to accessories that are no longer available.
Action taken
Valeant Pharmaceuticals International, Inc. sent Field Correction Notifications /Recall Stock Response Form dated September 22, 2015 through 2nd day air service to their customers. The notification informed the customer about the field correction, how to identify affected product and what actions to take. T The customers were requested to fill out and return the attached Recall Stock Response Form and return the form via email (bauschandlomb6443@stericycle.com or fax (844.782.5565). A customer service number was provided in case the customer had any related questions or concerns (1.800.321.4576) On October 7, 2015 a second field correction notification was sent to customers. The notification informed the customer about the field correction, how to identify affected product and what actions to take. A customer service number was provided in case the customer had any related questions or concerns (1.800.321.4576). The updated notification clarified that the notification applies to all SPAG-2 6000 model units currently in the market.
Root cause
Finished device change control
Status
Terminated
Product code
CAF
Quantity affected
1178
Distribution
USA (nationwide) Distribution and Internationally to Canada and France.
Date initiated
2015-09-21
Date posted
2016-01-21
Date terminated
2017-05-19
Location
Bridgewater, NJ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SPAG-2 AEROSOL GENERATOR W/USE OF VIRAZOLE (K870051)Viratek, Inc.1987-10-06
(SPAG) MODEL SPAG-2 6000 SERIES (K871427)Viratek, Inc.1987-10-06