Vacuum Collection Bottle, 1 gallon (3200 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0302-600 (6700-20
FDA device recall Z-0090-2010 · posted 2009-10-29 · Terminated
Recall details
- Recalling firm
- Ohio Medical Corporation
- Reason for recall
- The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on polycarbonate.
- Action taken
- Ohio Medical Corporation issued recall letters beginning September 17, 2009. Consignees were informed of the affected product and requested to return affected collection bottles to the firm. The accounts were requested to complete and return the enclosed acknowledgement form, indicating the number of affected units on hand, to obtain a returned material authorization for the return of the recalled bottles and their replacement. For further information, contact Ohio Medical Quality Assurance Manager, Anita Christ, at 847-855-6270.
- Root cause
- Process control
- Status
- Terminated
- Product code
- KDQ
- Quantity affected
- 823 bottles
- Distribution
- Worldwide Distribution -- United States, Canada, Chile, Costa Rica, Cyprus, Egypt, El Salvador, Greece, Hong Kong, India, Israel, Japan, Jordan, Korea, Kuwait, Malaysia, Netherlands, Oman, Paraguay, P
- Date initiated
- 2009-09-21
- Date posted
- 2009-10-29
- Date terminated
- 2011-06-22
- Location
- Gurnee, IL