Atlas FDA

Vacuum Collection Bottle, 1/2 gallon (1300 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0301-600 (6700-

FDA device recall Z-0089-2010 · posted 2009-10-29 · Terminated

Recall details

Recalling firm
Ohio Medical Corporation
Reason for recall
The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on polycarbonate.
Action taken
Ohio Medical Corporation issued recall letters beginning September 17, 2009. Consignees were informed of the affected product and requested to return affected collection bottles to the firm. The accounts were requested to complete and return the enclosed acknowledgement form, indicating the number of affected units on hand, to obtain a returned material authorization for the return of the recalled bottles and their replacement. For further information, contact Ohio Medical Quality Assurance Manager, Anita Christ, at 847-855-6270.
Root cause
Process control
Status
Terminated
Product code
KDQ
Quantity affected
5,892 bottles
Distribution
Worldwide Distribution -- United States, Canada, Chile, Costa Rica, Cyprus, Egypt, El Salvador, Greece, Hong Kong, India, Israel, Japan, Jordan, Korea, Kuwait, Malaysia, Netherlands, Oman, Paraguay, P
Date initiated
2009-09-21
Date posted
2009-10-29
Date terminated
2011-06-22
Location
Gurnee, IL

Monitor recalls programmatically