Uritek TC-101Urine Analyzer, catalog number TC-101

FDA device recall Z-0472-06 · posted 2006-02-04 · Terminated

Recall details

Recalling firm
Teco Diagnostics
Reason for recall
When reading the Nitrite parameter on the Teco Diagnostics URS-10 strips, some positive nitrite samples may be read by the machine as negative.
Action taken
On December 22nd,2005 the firm sent the notification letter along with a replacement unit of TC-101. These letters and replacement units were send via Fed ex ground and the tracking numbers and serial numbers of the sent units are also noted. Also noted are the return authorization numbers assigned to each respective customer.
Root cause
Other
Status
Terminated
Product code
KQO
Quantity affected
50
Distribution
Nationwide
Date initiated
2005-12-22
Date posted
2006-02-04
Date terminated
2012-03-07
Location
Anaheim, CA