Atlas FDA

***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgical environment

FDA device recall Z-0196-2024 · posted 2023-10-27 · Open, Classified

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc
Reason for recall
Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure
Action taken
Siemens Medical Solutions USA issued Urgent Medical Device Correction Letter (AX035/23/S).Letter states reason for recall, health risk and action to take. Siemens will correct the software error via Update Instruction AX036/23/S. With software version VA30K, the user interface of Cios VA30 systems will be changed to consider the current standards DIN 6862-3 and IEC 60601-2-54. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. Please respond within 14 days of receiving this notification. If you have received this notification by email via Adobe GigaSign, kindly sign with your digital signature to acknowledge that you have read and understand the content provided. If you have received this customer notification by delivery by the US Postal Service or by FedEx, please send an email to recallsandrefusals.team@siemens-healthineers.com acknowledging that you have read and understand the content provided. Please ensure that all users of the affected products within your organization and others who may need to be informed will receive the relevant safety information provided within this notice and will comply with the recommendations therein.
Root cause
Software design
Status
Open, Classified
Product code
OWB
Quantity affected
***Updated 12/4/23*** 186 units Cios Flow, 181 units Cios Spin
Distribution
Nationwide
Date initiated
2023-09-12
Date posted
2023-10-27
Location
Malvern, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Cios Spin (K210054)Siemens Medical Systems USA, Inc.2021-02-05
Cios Flow (K203504)Siemens Medical Solutions USA, Inc.2020-12-22