UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B68304

FDA device recall Z-1013-2017 · posted 2017-01-11 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. This is due to limitations in the available technology as well as sample limitations.
Action taken
Beckman Coulter sent an Urgent Medical Device Recall letter dated December 19, 2016, to all affected customers to inform them that Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. The letter informs the customers of the impact, actions to be taken, and resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Customer Support Center: http://www.beckmancoulter.com/customersupport/support, 800-526-7694 in the United States and Canada, and outside the United States and Canada, to contact their local Beckman Coulter Representative.
Root cause
Under Investigation by firm
Status
Terminated
Product code
GKZ
Quantity affected
4,657 units total (2,067 units in US)
Distribution
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the countries of : Australia, Belarus, Belgium, Bosnia and Herzegovina, Brunei Darussalam, Canada, China, Columbia,
Date initiated
2016-12-19
Date posted
2017-01-11
Date terminated
2017-05-12
Location
Brea, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM/ UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM (K140911)Beckman Coulter, Inc.2014-09-05
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM (K120771)Beckman Coulter, Inc.2013-03-22
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM (K081930)Beckman Coulter, Inc.2008-12-19