Unicel Closed Tube Aliquotter (UCTA) for UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, UniCel DxC 880i Synchron Acc
FDA device recall Z-0059-2013 · posted 2012-10-12 · Terminated
Recall details
- Recalling firm
- Beckman Coulter Inc.
- Reason for recall
- The recall was initiated because Beckman Coulter has confirmed that the current sample syringe pump drive assemblies installed on DxC Integrated ystem with UniCel Closted Tube Aliquotter (UCTA) serial numbers between 547 and 574 may cause a hardware motion error (ID0x08000501).
- Action taken
- The recall notification was initiated on July 6, 2012 with Beckman Coulter forwarding a Product Corrective Action (PCA) letter with attached PCA Response Form to all customers who purchased the UniCel DxC 880i, UniCel DxC 860i, UniCel 680i, UniCel DxC 660i Clinical System Integrated Workstations UCTA Sample Syringe Pump Drive Assembly. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were notified that Beckman Coulter will schedule a Service Call to replace the affected sample syringe pump drive assemblies. Customers were instructed to complete and return the enclosed PCA Response Form within 10 days. Customers with any questions regarding the PCA, were instructed to call Customer Support at (800) 854-3633 in the US and Canada. Outside the US and Canada, customers were instructed to contact Beckman Coulter Representative.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- JJE
- Quantity affected
- 18 units
- Distribution
- Worldwide Distribution and US (nationwide) and the countries of China, France, Italy and Taiwan.
- Date initiated
- 2012-07-06
- Date posted
- 2012-10-12
- Date terminated
- 2012-11-15
- Location
- Brea, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| UNICEL DXC 600 SYNCHRON SYSTEM SOFTWARE VERSION 5.0 AND UNICEL DXC 800 SYNCHRON SYSTEM SOFTWARE VERSION 5.0 (K103842) | Beckman Coulter, Inc. | 2011-04-27 |
| UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM (K060256) | Beckman Coulter, Inc. | 2006-02-27 |
| UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS (K042291) | Beckman Coulter, Inc. | 2004-11-12 |