Ultraview Spacelabs Medical Multigas Analyzer, Model 91518, V1.00.14. Spacelabs Healthcare Llc. Monitors gas concentrati

FDA device recall Z-2902-2011 · posted 2011-08-01 · Terminated

Recall details

Recalling firm
Spacelabs Healthcare, Llc
Reason for recall
Spacelabs received a report of the Model 91518 Multigas Analyzer trending low levels of an anesthetic agent that is not present during a particular case. The error is limited to only the trend table data.
Action taken
Spacelabs Healthcare, Llc. sent a "Customer Bulletin" dated June 20, 2011 to all affected customers. The customers were notified of the trend problem and are advised to follow the instructions in the Perioperative section and the ADMIT/DISCHARGE section of the users manual. For questions contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support or 1-425-657-7200 x5089.
Root cause
Software design
Status
Terminated
Product code
CBR
Quantity affected
481 units worldwide (333 units internationally and 148 units in the US)
Distribution
Worldwide Distribution: (USA) Nationwide distribution, and the countries of ARGENTINA, BAHRAIN, BOLIVIA, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, ECUADOR, EGYPT, FRANCE, INDIA, MEXICO, NICARAGU
Date initiated
2011-06-08
Date posted
2011-08-01
Date terminated
2011-09-21
Location
Issaquah, WA