UltraLite Vehicles powered by Rascal Motorized Vehicles.
FDA device recall Z-1376-05 · posted 2005-08-17 · Terminated
Recall details
- Recalling firm
- Electric Mobility Corp
- Reason for recall
- The engager set screw may become loose. This problem may cause the potentiometer to become off-center thus affecting the end users stopping capabilities.
- Action taken
- The firm sent out letters to their customers on 4/20/2005. The firm's sales representatives were also notified on 4/20/2005.
- Root cause
- Other
- Status
- Terminated
- Product code
- ITI
- Quantity affected
- 788 units
- Distribution
- The firm sells the vehicles via direct sales to end customers and also to company owned inventory with sales representatives. The distribution is nationwide.
- Date initiated
- 2005-05-05
- Date posted
- 2005-08-17
- Date terminated
- 2006-10-03
- Location
- Sewell, NJ