Tucker Embryo Catheter with URM Tip, Embryo Transfer Catheter, 23 cm, Order #: ftr123, Manufactured by: Fertility Techno
FDA device recall Z-0665-2008 · posted 2008-01-19 · Terminated
Recall details
- Recalling firm
- Fertility Technology Resources, Inc.
- Reason for recall
- Expiration Date for sterility; The 5 year dating could not be validated/supported for sterility assurance.
- Action taken
- All customers were contacted by telephone starting 9/13/2007. They were asked to return all Tucker family of catheters and all Morton IUI catheters that are on the shelves. A UPS call tag was sent to each customer to facilitate the return of the products. A follow up Urgent Device Recall letter was sent to all customers on 10/9/2007 reminding them to return all outstanding inventory when the UPS call tag arrives.
- Root cause
- Process design
- Status
- Terminated
- Product code
- MQF
- Quantity affected
- not reported
- Distribution
- Worldwide: USA including states of CA, CO, FL, GA, LA, MD, ME, MI, MS, NC, NH, NJ, NY, OH, TN, and WA and The Netherlands.
- Date initiated
- 2007-09-13
- Date posted
- 2008-01-19
- Date terminated
- 2008-02-06
- Location
- Marietta, GA