Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 6

FDA device recall Z-0816-05 · posted 2005-05-11 · Terminated

Recall details

Recalling firm
Del Medical Systems Group
Reason for recall
The Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator.
Action taken
Del Medical sent Field Correction Recall letters dated 3/10/05 to the direct accounts, informing them of the potential for a defective weld supporting the braket holding the locking cylinder assembly. If the weld fails, the tubehead arm assembly could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.
Root cause
Other
Status
Terminated
Product code
IYB
Quantity affected
31 stands
Distribution
Nationwide, Mexico and Thailand.
Date initiated
2005-03-10
Date posted
2005-05-11
Date terminated
2005-11-14
Location
Franklin Park, IL