TSX-301A/2 Aquilion One CT System. The TSX-30IAl2 is a whole body CT scanner. This device captures cross sectional volum

FDA device recall Z-2468-2012 · posted 2012-09-27 · Terminated

Recall details

Recalling firm
Toshiba American Medical Systems Inc
Reason for recall
The SureStart function may not operate in the usual manner, and you may obtain abnormal analysis results when using the cerebral blood flow analysis function.
Action taken
Toshiba American Medical Systems, Inc. sent an "URGENT: MEDICAL DEVICE CORRECTION" letter to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. customers: The firm advised customers to take the following temporary measures to avoid issues until a corrective action has been taken: I. In SureStart: " Perform scanning with Observation Direction set to VFF when the Patient Insertion Direction and Observation direction are FF. " Perform scanning with the Observation direction set to VFH when the Patient Insertion Direction and Observation Direction are HF. 2. When acquiring images for 4D perfusion analysis: " Perform scanning with the Scan Direction set to OUT when the Patient Insertion Direction and Scan Direction are HF. Perform scanning with the scan direction set to IN when the Patient Insertion Direction and Scan Direction are FF. Contact the firm at (800) 421-1968 for questions regarding this issue.
Root cause
Software design
Status
Terminated
Product code
JAK
Quantity affected
3 units
Distribution
Nationwide distribution: USA including states of: MA, MD, and NV.
Date initiated
2008-05-13
Date posted
2012-09-27
Date terminated
2012-09-27
Location
Tustin, CA