Trypticase TM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar Cat. Number 221291, Lot Number 6034889, Exp.
FDA device recall Z-1559-2016 · posted 2016-04-22 · Terminated
Recall details
- Recalling firm
- Becton Dickinson & Co.
- Reason for recall
- Trypticase TM Soy Agar with 5% Sheep Blood (TSA IITM) // MacConkey II Agar is contaminated with Listeria Monocytogenes. The device is a non-sterile prepared plated media.
- Action taken
- BD sent an Urgent Product Recall letter dated March 2016 to all US customers via UPS or email. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the affected lots and discard any remaining product. BD will replace the discard inventory. Customers were asked to complete and fax the attached form to : Attention Regulatory Compliance, Fax: 410-316-4258 or email to BDRC2@bd.com.
- Root cause
- Material/Component Contamination
- Status
- Terminated
- Product code
- JSI
- Quantity affected
- 88,300 plates (883 cartons)
- Distribution
- US Nationwide Distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, KS, LA, MA, MD, ME, MI, MN, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TX, UT, VA, WA.
- Date initiated
- 2016-03-28
- Date posted
- 2016-04-22
- Date terminated
- 2016-11-14
- Location
- Sparks, MD
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.