Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the followin

FDA device recall Z-0139-2025 · posted 2024-10-22 · Open, Classified

Recall details

Recalling firm
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Reason for recall
If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.
Action taken
On 9/6/24, recall notices were sent to customers who were asked to do the following: 1) Firmware for the system controller requires updating. The work to correct this issue will be performed by your Authorized Service provider. 2) Schedule time with your Distributer to have your system updated. 3) Complete and return the response form to drodriguez@shimadzu-usa.com Questions can be directed to the firm's National Technical Support: Toll Free: (844)487-2767 Available 8:00 to 5:00 PM, Monday - Friday
Root cause
Software Design Change
Status
Open, Classified
Product code
OWB
Quantity affected
13
Distribution
US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.
Date initiated
2024-09-06
Date posted
2024-10-22
Location
Kyoto, Japan

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
TRINIAS (K123508)Shimadzu Corporation2014-02-26