Triglycerides GPO. In vitro test for the quantitative determination of triglycerides in human serum and plasma on the CO
FDA device recall Z-1929-2015 · posted 2015-06-27 · Terminated
Recall details
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Reason for recall
- Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
- Action taken
- Roche Diagnostics initiated a voluntary recall and the Urgent Medical Device Correction notices were sent via UPS Ground on 05/07/2015 to customers. Customers were notified of the interference and the method sheets and package inserts will be updated with the new interference claims. The firm plans to monitor any non-responding accounts and address those customers with follow-up attempts to distribute recall notices. The firm plans to conduct level B effectiveness checks via fax and phone. Until the new method sheets are available, the Urgent Medical Device Correction serves as the updated labeling. Roche is not requesting return of the affected product. Customers may contact Todd Siesky Corporate Communications Director, Phone: 317.521.3911, Fax: 317.521.4696, E-mail: todd.siesky@roche.com. For technical questions, contact the Customer Support Center at 1-800-428-2336
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- JJE
- Quantity affected
- 6255
- Distribution
- US (nationwide) including PR.
- Date initiated
- 2015-05-07
- Date posted
- 2015-06-27
- Date terminated
- 2017-03-24
- Location
- Indianapolis, IN
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| BOEHRINGER MANNHEIM/ HITACHI 911 ANALYZER (K921661) | Boehringer Mannheim Corp. | 1992-06-30 |