Atlas FDA

Triathlon X3 Total Stabilizer + Tibial Insert For use with Triathlon Universal Baseplates, knee prosthesis. #5 13mm Stry

FDA device recall Z-1231-2007 · posted 2007-09-25 · Terminated

Recall details

Recalling firm
Stryker Howmedica Osteonics Corp.
Reason for recall
Stabilizer posts, packaged with the Triathlon TS Plus Tibial inserts have a specification anomaly related to length, resulting in a potential for the post to slightly protrude above the top surface of the plastic post.
Action taken
Recall notification letters and product acknowledgment forms were send via Federal Express on August 2, 2007.
Root cause
Other
Status
Terminated
Product code
JWH
Quantity affected
14
Distribution
Product distributed to 5 hospitals in GA, OH, MN, TX.
Date initiated
2007-07-31
Date posted
2007-09-25
Date terminated
2007-12-05
Location
Mahwah, NJ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
TRIATHLON TS KNEE SYSTEM (K070095)Howmedica Osteonics Corp.2007-06-01