Triathlon PKR Peg Drill, Catalog Number: 5650-4-512 (Size 1-2 Peg Drill); Catalog Number: 5650-4-536 (Size 3-6 Peg Drill
FDA device recall Z-1835-2011 · posted 2011-03-28 · Terminated
Recall details
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Reason for recall
- Stryker Orthopaedics has identified a trend of complaints involving reports of the peg drill cold welding to the drill guide, causing the Peg Drill to cease functioning. If cold welding of these components occurs, the following potential patient and/or surgeon or surgical staff harms may results.
- Action taken
- Stryker sent Urgent Product Recall letters on February 9, 2011, and February 10, 2011, via Fed Ex. to all affected customers. The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to complete and fax back the attached Product Recall Acknowledgement Form as soon as possible to 201-831-6069. For questions regarding this recall call 201-972-2100.
- Root cause
- Component change control
- Status
- Terminated
- Quantity affected
- 5650-4-512- 754 units; 5650-4-536- 880 units
- Distribution
- Worldwide Distribution - USA , Sweden, Germany; France, Portugal, UK, Canada, Australia, and New Zealand,
- Date initiated
- 2011-02-09
- Date posted
- 2011-03-28
- Date terminated
- 2013-11-18
- Location
- Mahwah, NJ