Triathlon MIS AP Sizer Body - Left and Right; Catalog Number: 6541-5-508- left; Catalog Number: 6541-5-509 - right, Stry
FDA device recall Z-0448-2008 · posted 2007-12-11 · Terminated
Recall details
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Reason for recall
- Assembly problems: MIS instruments (Triathlon MIS AP Sizer Body - Left; Triathlon MIS Sizer Body - Right; Triathlon MIS Femoral Adjustment Block and Triathlon MIS Modular Capture) may not assemble properly.
- Action taken
- Important Product Correction letters and Product Accountability Forms were sent via Federal Express on 9/19/06 with return receipt. Foreign locations were notified by a Advisory Notice. Notification letters requested that all inventory of the affected items be examined to verify that it assemble correctly. Any product that does not assemble correctly was to be returned to Stryker, Mahwah, NJ. The Product Accountability form was to be completed by each location to notify Stryker that the letter ahd been received and inventory examined. Stryker representatives were to be contacted for product replacement.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- JWH
- Quantity affected
- 2,868 units total for Recalls Z-0448-0450-2008
- Distribution
- Worldwide: USA including states of MI and IL. and countries of Australia, China, New Zealand, Singapore, France, Germany, Italy, Portugal, Spain, Sweden and UK.
- Date initiated
- 2006-09-19
- Date posted
- 2007-12-11
- Date terminated
- 2008-01-18
- Location
- Mahwah, NJ