Toshiba Infinix-i X-Ray System. A radiographic and fluoroscopic x-ray system for use in diagnostic and angiographic proc

FDA device recall Z-0994-2013 · posted 2013-03-25 · Terminated

Recall details

Recalling firm
Toshiba American Medical Systems Inc
Reason for recall
Lost and incorrect images. When using the "F-Rec" fluoroscopy image acquisition recording option in the "manual mode" an infrequent software error may cause images from another study to be displayed on the monitor. The acquired images may be lost requiring additional image acquisitions to be performed.
Action taken
In November, 2012, Toshiba America Medical Systems, Inc. sent an Urgent: Medical Device Correction letter to customers. The letter contains a return reply form that is to be faxed to TAMS for retention.
Root cause
Software design
Status
Terminated
Product code
MQB
Quantity affected
95 units
Distribution
USA nationwide. No international.
Date initiated
2012-11-08
Date posted
2013-03-25
Date terminated
2013-03-26
Location
Tustin, CA