Toshiba Excelart VANTAGE Model MRT-1503, Magnetic Resonance Imaging Systems, Regulation Number 21 CPR 892.1000, Regulato

FDA device recall Z-0281-06 · posted 2005-12-14 · Terminated

Recall details

Recalling firm
Toshiba American Med Sys Inc
Reason for recall
To correct a hardware problem of water leakage in the cooling system inside the ASGC, leading to the replacement of the gradient coil.
Action taken
Toshiba America Medical Systems, Inc., (TAMS) customer service engineers will hand deliver The Advisory Letter and monitor the system during their monthly preventive maintenance.
Root cause
Other
Status
Terminated
Product code
LNH
Quantity affected
65
Distribution
Nationwide
Date initiated
2005-10-12
Date posted
2005-12-14
Date terminated
2007-07-12
Location
Tustin, CA