Toshiba Excelart VANTAGE Model MRT-1503, Magnetic Resonance Imaging Systems, Regulation Number 21 CPR 892.1000, Regulato
FDA device recall Z-0281-06 · posted 2005-12-14 · Terminated
Recall details
- Recalling firm
- Toshiba American Med Sys Inc
- Reason for recall
- To correct a hardware problem of water leakage in the cooling system inside the ASGC, leading to the replacement of the gradient coil.
- Action taken
- Toshiba America Medical Systems, Inc., (TAMS) customer service engineers will hand deliver The Advisory Letter and monitor the system during their monthly preventive maintenance.
- Root cause
- Other
- Status
- Terminated
- Product code
- LNH
- Quantity affected
- 65
- Distribution
- Nationwide
- Date initiated
- 2005-10-12
- Date posted
- 2005-12-14
- Date terminated
- 2007-07-12
- Location
- Tustin, CA