TomoTherapy Hi-Art System, Version 3.X ThomoTherapy Incorporated 1240 Deming Way, Madison, WI 53717. The TomoTherapy HI-
FDA device recall Z-2243-2009 · posted 2009-09-21 · Terminated
Recall details
- Recalling firm
- TomoTherapy Incorporated
- Reason for recall
- TomoTherapy received one customer report indicating that after starting a patient procedure from the Status Console, the radiation on light and the audible indicator on the Status Console came on as expected, however the Operator Station (OS) did not display the dose rate and the procedure time, or the couch position updates. Within a few minutes, the customer pressed the Stop button on the Sta
- Action taken
- Consignees were sent a TomoTherapy "Urgent Medical Device Correction Field Safety Notice" letter dated August 10, 2009. The letter addressed the Issue, Product Affected, Investigation, Recommended Actions and Resolution. Customers were recommended to perform the following actions (in the event that the dose rate, procedure timer, and couch position updates are not present on the OS, after pressing Start on the status console and the radiation on light and audible indicators are active): 1) Immediately press the Stop button on the Status Console and 2) contact your authorized TomoTherapy Service Representative for further direction. The affected issue will be corrected in the next software version, which is scheduled for release in early 2010. The TomoTherapy Customer Contact Center can be reached 24 hours a day, 7 days a week, by e-mail at support@tomotherapy.com or by phone at 1-866-368-4807 (North America).
- Root cause
- Software design
- Status
- Terminated
- Product code
- MUJ
- Quantity affected
- 259 (169 US; 91 OUS)
- Distribution
- Worldwide Distribution -- US (states of AL, AK, AZ, AR, CA, CO, CT, DC, FL, GA, HI, IL IN, IA, KS, KY, LA, MD, MA, MI, MN,MO, NE,NV,NJ, NM, NY, OH, OK OR, PA, PR, SC, SC, TN, TX, UT, VA, WA, WI and WY
- Date initiated
- 2009-08-10
- Date posted
- 2009-09-21
- Date terminated
- 2011-12-17
- Location
- Madison, WI