FDA device recall Z-0773-2012 · posted 2012-01-17 · Terminated

Recall details

Recalling firm
Microgenics Corp
Reason for recall
Reports that third party control material for 4 lots of Tobramycin Calibrators that are recovering low and outside of the published range.
Action taken
ThermoFisher Scientific sent a product removal letter dated November 11, 2011. Letters were sent to the sole distributor of the product. The letter identified the product the problem and the action needed to be taken by the customer. If you are using or have inventory of the affected lot numbers immediately discontinue use and destroy any remaining affected lot numbers per your local waste ordinances. If you have any further questions regarding this information please call 1-877-4Abbott. Customers outside the US should contact their local customer service.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DLJ
Quantity affected
1366 kits
Distribution
Worldwide Distribution -- Nationwide Distribution including the country Europe.
Date initiated
2011-11-11
Date posted
2012-01-17
Date terminated
2012-05-09
Location
Fremont, CA