Tobii C-Series Lithium batteries Type TB8A, TB8B, TB12A used with the Tobii C8 and Tobii C12 Speech Generating Communica

FDA device recall Z-2694-2011 · posted 2011-06-29 · Terminated

Recall details

Recalling firm
Tobii Assistive Technology, Inc.
Reason for recall
Lithium Batteries may overheat and become combustible
Action taken
Tobii Technology AB sent an "URGENT: MEDICAL DEVICE RECALL" letter dated May 25, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were advised to stop using the batteries and to remove them from their devices. They were instructed to dispose of the batteries in an environmentally responsible manner as hazardous materials. Tobii will replace the affected batteries. Customers were asked to complete and return an acknowledgement form via fax at 781-461-8213 or e-mail at recall@tobiiati.com. Contact the firm by E-mail: recall@tobiiati.com or telephone: 1-800-793-9227.
Root cause
Environmental control
Status
Terminated
Product code
ILQ
Quantity affected
2430 units
Distribution
USA (nationwide)
Date initiated
2011-05-31
Date posted
2011-06-29
Date terminated
2012-10-01
Location
Dedham, MA