TiTLE 2 Polyaxial Spinal System, Counter Torque Assembly Tool, Part ET1132-01, Zimmer, Warsaw, IN. The Zimmer Title 2 co

FDA device recall Z-0887-2011 · posted 2011-01-10 · Terminated

Recall details

Recalling firm
Zimmer Inc.
Reason for recall
There is a chance that the ball plunger could fall out during surgery and fall into the surgical site. There could be a delay in surgery to remove the ball plunger. If the problem was undetected, the ball plunger could remain in the patient causing pain and potentially necessitating subsequent surgery for removal.
Action taken
Zimmer sent an URGENT MEDICAL DEVICE RECALL LETTER dated October 21, 2010, to all affected customers. The letter identified the product, the problem, the action to be taken by the customer. Customers were instructed to: Inspect their inventory, stop using the device, and place it in quarantine. Contact Zimmer Customer Service at 1-800-774-6368 to obtain a return authorization number. Complete the Inventory Return Certification form and fax a copy to (952) 837-6958, or scan and email to qualitycompliance@zimmer.com. Return the recalled product along with the original completed Return Certification Form to: Zimmer Spine, Inc. Attn: RA Receiving 5301 Riata Park Ct. Austin, Texas 78727 For questions regarding this recall call Zimmer Spine Customer Service at 1-866-774-6368. .
Root cause
Other
Status
Terminated
Product code
LXH
Quantity affected
68
Distribution
Nationwide Distribution including AL, AZ, CA, CO, FL, MI, MN, NC, PA, OH, OK, TN, TX, WA, and WI.
Date initiated
2010-10-21
Date posted
2011-01-10
Date terminated
2011-10-12
Location
Warsaw, IN