Titan T 5K Motor Product Usage: Low-speed motor is used with various attachments to perform a variety of dental procedur

FDA device recall Z-2078-2014 · posted 2014-07-17 · Terminated

Recall details

Recalling firm
Dental EZ Group Star Dental Division
Reason for recall
The housing of 5K motor was laser marked incorrectly as a 20K motor. The recall was expanded in September 2014 to include additional lot #2251181.
Action taken
Star Dental sent a Medical Device Recall letter, dated June 13, 2014 to affected customers. This letter identified the affected product, problem, and actions to be taken. For question call 866-383-4636 extension 4350. The recall was expanded in September 2014 to include additional lot #2251181. A letter, dated September 2014 was sent to customers who were impacted by the recall expansion.
Root cause
Process control
Status
Terminated
Product code
EFA
Quantity affected
45
Distribution
US Distribution in the states of : IN, TN, WA, CA, UT, MN, and NY.
Date initiated
2014-06-13
Date posted
2014-07-17
Date terminated
2015-02-24
Location
Lancaster, PA