Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment
FDA device recall Z-1232-2017 · posted 2017-02-22 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.
- Action taken
- A recall notification is being forwarded to all users and consignees who have received product. The notification of this voluntary recall will be sent to the following consignee in each facility:+ Director of Operating Room Services Proposed Level of Effectiveness Checks: Terumo CVS will verify that 100% of all letters are delivered to affected consignees
- Root cause
- Other
- Status
- Terminated
- Product code
- MWS
- Quantity affected
- 1463
- Distribution
- The product was distributed domestically and also to foreign consignees. But not to government agency's or Canada.
- Date initiated
- 2015-01-05
- Date posted
- 2017-02-22
- Date terminated
- 2017-02-23
- Location
- Elkton, MD