Tina-Quant (Latex) HS Test System (C-Reactive Protein (Latex) High Sensitive) The Tina-Quant CRP (Latex) High Sensitive
FDA device recall Z-1092-2017 · posted 2017-01-24 · Terminated
Recall details
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Reason for recall
- Tina-quant Cardiac C-reactive Protein (Latex) High Sensitive (CRPHS) lot number 604450 (catalog number 11972855216) shows an under-recovery of up to -25% at concentrations below 5 mg/L on the MODULAR ANALYTICS P module. this issue can lead to erroneous low CRPHS results. CRPHS is used as a marker for the prediction of cardiovascular risk. An erroneous low result could lead to an incorrect cardiovascular risk assessment and subsequent delay of further examination.
- Action taken
- Roche Diagnostics sent a Reagent Bulletin dated February 26, 2016, to all affected consignees. The recall notifications included a description of the reason for the recall, affected product, request for replacement, and consignee responsibilities; the notification did not include instructions for responding to the recall notification. Actions Required " Do not use CRPHS lot number 604450. If you have any affected product in your inventory, discard it per local guidelines. " Complete the attached replacement request form to receive replacement product. " File this Reagent Bulletin for future reference. Questions Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this Reagent Bulletin.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- NQD
- Quantity affected
- 3204
- Distribution
- Nationwide Distribution including AL, AZ, CA, CT, FL, GA, KS, MA, MO, NJ, NY, PA, and WA.
- Date initiated
- 2016-02-26
- Date posted
- 2017-01-24
- Date terminated
- 2017-06-05
- Location
- Indianapolis, IN