Atlas FDA

Tiger Screw Sterilization Tray Base Assembly Model 210-00-001 Revision N, O The product is a component of a system indic

FDA device recall Z-0960-2017 · posted 2017-01-06 · Terminated

Recall details

Recalling firm
Trilliant Surgical Ltd.
Reason for recall
Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.
Action taken
Trilliant Surgical Ltd. sent an Urgent Medical Device Recall letter dated December 8, 2016, to all affected consignees. The letter instructed the consignees to 1) identify any/all affected product; 2) document any/all identified product via a provided response form and return to Trilliant Surgical; 3) exchange product upon receipt of replacements; and 4) return any/all affected product as originally identified and reported on the return response form. Customers with questions were instructed to call 800-495-2919.
Root cause
Under Investigation by firm
Status
Terminated
Product code
HWC
Quantity affected
50 units
Distribution
Nationwide Distribution
Date initiated
2016-12-08
Date posted
2017-01-06
Date terminated
2017-05-17
Location
Houston, TX