Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the
FDA device recall Z-1281-2015 · posted 2015-03-16 · Terminated
Recall details
- Recalling firm
- Synthes, Inc.
- Reason for recall
- For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
- Action taken
- An Urgent Notice: Medical Device Recall, dated February 17, 2015, was sent to all direct consignees and sales consultants instructing them to take the following actions: Take note of the quantity of affected parts in your possession, and then follow ONE of the options for replacement of the affected product: Option 1: Call DePuy Synthes Repair Department at 1-800-288-6698 to schedule a repair for the affected products. Complete the Verification Section following the instructions provided and send a copy of the completed Verification Section to DePuy Synthes by: Fax: 866-723-2823 or Scan/e-mail: Synthes6734@stericycle.com Option 2: Call DePuy Synthes Repair Department at 1-800-288-6698 to order repair part. The replacement part will be sent directly to your facility. Questions should be directed to 610-719-5450 or a local DePuy Synthes sales consultant.
- Root cause
- Process control
- Status
- Terminated
- Product code
- FZX
- Quantity affected
- 183
- Distribution
- Nationwide Distribution
- Date initiated
- 2015-02-17
- Date posted
- 2015-03-16
- Date terminated
- 2016-08-26
- Location
- West Chester, PA