This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative
FDA device recall Z-0823-2016 · posted 2016-02-19 · Terminated
Recall details
- Recalling firm
- Leica Microsystems, Inc.
- Reason for recall
- The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining when using a detection protocol where one of its components - Peroxide Block / Peroxidase Block is applied after the primary marker incubation as specified in the Instructions for Use Following the Instructions for Use.
- Action taken
- The firm sent out a Medical Device Recall letter dated December 29th 2015 to two-hundred (200) consignees in the United States decribing the recall and the neccesary actions to be taken. These included: - Do not use or continue to use the listed antibody with the specified detection kits. - Order replacement detection kits for use with the listed antibody. - There is no need for retrospective review of tissue staining if appropriate positive controls were properly used and interpreted during the immuno / in-situ staining. -pass this notice primarily to the end users where the product has been sold and to all those within your organization who need to be aware of this issue. -Confirm reciept of the Recall Notice letter by signing and dating the attached Acknowledgement Form and faxing it back to us at 1-847-236-3747, or scan and email it to LMGRA@Leica-Microsystems.com Customer are requested to contact their local Leica representative if you have any questions or concerns, or alternatively please contact us at the following email address: Combination.Products@LeicaBiosystems.com Peter Lloyd Leica Biosystems Newcastle Ltd Balliol Business Park West Benton Lane Newcastle upon Tyne NE 12 8EW United Kingdom Telephone +44 191 215 0567 Facsimile: +44 191 215 1152
- Root cause
- Component change control
- Status
- Terminated
- Product code
- NJT
- Quantity affected
- 1700 units
- Distribution
- US Consignees incluude: AR ,AZ ,CA ,CO ,CT ,DC ,DE ,FL ,GA ,HI ,IA ,IL ,KY ,LA ,MA ,MD ,MI ,MN ,MO ,MT ,NC ,NH ,NJ ,NY ,OH ,OK ,PA ,PR ,SC ,TN ,TX ,UT ,VA ,VT ,WA ,WI and WV. Foreign Consignees includ
- Date initiated
- 2015-12-30
- Date posted
- 2016-02-19
- Date terminated
- 2017-01-05
- Location
- Buffalo Grove, IL